RFDA LTVP800
DESCRIPTION
The LTVP800 system is designed to perform impulse excitation measurements at temperatures from -50 °C up to 800 °C in vacuum (10-4 mbar). Measurements can be performed in predefined intervals during heating and cooling (1-60 °C/min) in order to determine the elastic properties and damping as function of the temperature. The LTVP800 system is equipped with a turbomolecular vacuum pump to perform measurements in a controlled vacuum atmosphere and a laser vibrometer to detect the vibration signal.
Internal dimensions of the furnace | Diameter: 350 mm Depth: 350 mm |
Temperature range | -50 °C – 800 °C |
Heating elements | Infrared heating |
Heating rate
|
1 – 60 °C/min
|
Isolation material
|
Vacuum |
Sample cooling
|
Indirectly by liquid nitrogen
|
Furnace wall cooling
|
Water jacket
|
Atmosphere | Vacuum |
Vacuum pump
|
Turbomolecular pump |
Minimum pressure
|
10-4 mbar
|
Furnace (sample) loading
|
Front load
|
Number of samples
|
1 |
Max. length of sample
|
100 mm |
Positioning excitation system
|
Fixed (1 position) |
Positioning microphone
|
Fixed (1 position) |
Laser vibrometer
|
Yes |